User Contributed MET/CAL PROCEDURE ============================================================================= INSTRUMENT: Honeywell UDC 3000 Type T °F - Low DATE: 10-Sep-97 AUTHOR: User Contributed REVISION: 0 ADJUSTMENT THRESHOLD: 70% NUMBER OF TESTS: 12 NUMBER OF LINES: 102 CONFIGURATION: Fluke 5700A STANDARD: Omega TRC III ============================================================================= STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON # This procedure is run with the MET/CAL Configuration #1 loaded. 1.001 ASK- R N P F W 1.002 ASK+ X 1.003 STD Omega TRC III 1.004 HEAD ADDITIONAL REQUIRED EQUIPMENT 1.005 DISP Omega TRC III Ice Point Cell 1.005 DISP Omega Type "T" Reference Thermocouple 1.006 HEAD INITIAL CONDITIONS 1.007 DISP Press UUT LOWR DISP button. 1.008 DISP Insert the reference thermocouple into the center 1.008 DISP well of the Ice Point Cell. 1.009 DISP Connect the copper end (white connector) to the 1.009 DISP 5700 OUTPUT HI/LO. Red wire is HI, green wire is LO. 1.010 DISP Connect a "T" type connector with "T" type T/C wire 1.010 DISP to UUT as follows (for UUT with short body use numbers 1.010 DISP in parenthesis): 1.010 DISP "T" probe TO UDC 3000 1.010 DISP OUTPUT + ................ terminal #17 (26) 1.010 DISP OUTPUT - ................ terminal #18 (27) 1.011 HEAD {* ThermoCouple Tests *} 1.012 HEAD { "T" Type T/C Low } 1.013 JMP 2.001 1.014 EVAL 2.001 5700 -4.149mV S 2W 2.002 MEMI Use keyboard to enter actual UUT reading in °Farenhite. 2.003 ACC -200.0°F 0.1U 2.004 MEME 2.005 MEMC °F 1.7U #! Test Tol 1.7, Sys Tol 0.1, TUR 17.0000 (>= 4.00). 3.001 5700 -2.581mV S 2W 3.002 MEMI Use keyboard to enter actual UUT reading in °Farenhite. 3.003 ACC -100.0°F 0.1U 3.004 MEME 3.005 MEMC °F 1.7U #! Test Tol 1.7, Sys Tol 0.1, TUR 17.0000 (>= 4.00). 4.001 5700 -0.674mV S 2W 4.002 MEMI Use keyboard to enter actual UUT reading in °Farenhite. 4.003 ACC 0.0°F 0.1U 4.004 MEME 4.005 MEMC °F 1.7U #! Test Tol 1.7, Sys Tol 0.1, TUR 17.0000 (>= 4.00). 5.001 5700 0mV S 2W 5.002 MEMI Use keyboard to enter actual UUT reading in °Farenhite. 5.003 ACC 32.0°F 0.1U 5.004 MEME 5.005 MEMC °F 1.7U #! Test Tol 1.7, Sys Tol 0.1, TUR 17.0000 (>= 4.00). 6.001 5700 1.518mV S 2W 6.002 MEMI Use keyboard to enter actual UUT reading in °Farenhite. 6.003 ACC 100.0°F 0.1U 6.004 MEME 6.005 MEMC °F 1.7U #! Test Tol 1.7, Sys Tol 0.1, TUR 17.0000 (>= 4.00). 7.001 5700 3.967mV S 2W 7.002 MEMI Use keyboard to enter actual UUT reading in °Farenhite. 7.003 ACC 200.0°F 0.1U 7.004 MEME 7.005 MEMC °F 1.7U #! Test Tol 1.7, Sys Tol 0.1, TUR 17.0000 (>= 4.00). 8.001 5700 6.647mV S 2W 8.002 MEMI Use keyboard to enter actual UUT reading in °Farenhite. 8.003 ACC 300.0°F 0.1U 8.004 MEME 8.005 MEMC °F 1.7U #! Test Tol 1.7, Sys Tol 0.1, TUR 17.0000 (>= 4.00). 9.001 5700 9.523mV S 2W 9.002 MEMI Use keyboard to enter actual UUT reading in °Farenhite. 9.003 ACC 400.0°F 0.1U 9.004 MEME 9.005 MEMC °F 1.7U #! Test Tol 1.7, Sys Tol 0.1, TUR 17.0000 (>= 4.00). 10.001 5700 12.572mV S 2W 10.002 MEMI Use keyboard to enter actual UUT reading in °Farenhite. 10.003 ACC 500.0°F 0.1U 10.004 MEME 10.005 MEMC °F 1.7U #! Test Tol 1.7, Sys Tol 0.1, TUR 17.0000 (>= 4.00). 11.001 5700 15.769mV S 2W 11.002 MEMI Use keyboard to enter actual UUT reading in °Farenhite. 11.003 ACC 600.0°F 0.1U 11.004 MEME 11.005 MEMC °F 1.7U #! Test Tol 1.7, Sys Tol 0.1, TUR 17.0000 (>= 4.00). 12.001 END #! T.U.R.s less than 4.00: 0 #! T.U.R.s estimated using RANGE value: 0 #! T.U.R.s not calculated (ASK- U): 0 #! T.U.R.s not computable at compile time: 0 #! FOR JUSTIFICATION REFER TO COMMENTS FOLLOWING EACH TEST IN THIS LISTING.